EasyManua.ls Logo

Rohde & Schwarz FSEA20 - Page 203

Rohde & Schwarz FSEA20
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
READ-Subsystem FSE
1065.6016.12 6.146 E-16
COMMAND PARAMETERS UNIT COMMENT
READ
:BURSt
:MACCuracy
:OSUPpress
:STATus?
:AVERage?
:MAXimum?
:PERCentile
:STATus?
:AVERage?
:MAXimum?
:FREQuency
:STATus?
:AVERage?
:MAXimum?
--
--
--
--
--
--
--
--
--
query only
query only
query only
query only
query only
query only
query only
query only
query only
:READ:BURSt:PERRor:RMS:STATus?
This command starts the measurement of the phase and frequency error of the base station or
mobile and reads out the status of the RMS-measurement of the phase error taken over the selected
number of bursts.
0: failed, 1: passed
Example: ":READ:BURS:PERR:RMS:STAT?"
Features: *RST value: --
SCPI: device-specific
Modes:BTS, MS
When the measurement is started the instrument automatically assumes the SINGLE mode.
An ongoing measurement can be aborted via the command ABORt. Further results of the
phase/frequency error measurement can be then queried without restart of the measurement via the
:FETCh:BURSt-subsystem.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the phase/frequency error is selected (see :CONFigure:BURSt:PFERror).
:READ:BURSt:PERRor:RMS:AVERage?
This command starts the measurement of the phase and frequency error of the base station or
mobile and reads out the average of the RMS-measurement of the phase error taken over the
selected number of bursts.
Example: ":READ:BURS:PERR:RMS:AVER?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
When the measurement is started the instrument automatically assumes the SINGLE mode.
An ongoing measurement can be aborted via the command ABORt. Further results of the
phase/frequency error measurement can be then queried without restart of the measurement via the
:FETCh:BURSt-subsystem.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the phase/frequency error is selected (see :CONFigure:BURSt:PFERror).

Table of Contents