READ-Subsystem FSE
1065.6016.12 6.150 E-16
Carrier Power Individual Messung: (:CONFigure:MS:POWer:SINGle:STATe ON)
This command starts the measurement of the maximum output power of the base station or mobile
and reads out the result. The power control level is preset ( command :CONFigure<1|2>[:MS]:
POWer:LEVel <num_value>)
Parameter: The result is read out as an ASCII string in the following format: single
measurements are retrieved:
<Static Power Ctrl>,<Dyn Power Ctrl>,<Rat-Level>,<Act-
Level>,<RBW>,<Arfcn>,<CF>,<Attenuation>,<Number of bursts>,<Status>
<Static Power Ctrl>: current static power control level
<Dyn Power Ctrl>: current dynamic power control levell
<Rat-Level>: rated value for the current power control level acc.
to standard in dBm
<Act-Level>: measured power in dBm
<RBW>: resolution bandwidth in kHz
<ARFCN>: channel number
<CF>: carrier frequencyin Hz
<Att>: external attenuation in dBm
<Number of burst>: number of bursts
<Status>: result of limit check in character data form:
PASSED no limits exceeded
FAILED limit exceeded
Example: ":READ:BURS:POW?"
Result:: 0,3,37,20.6915,1000,2,8.904E+008,20,1,FAILED,
0,3,37,20.3597,1000,2,8.904E+008,20,1,FAILED
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
When the measurement is started any ongoing measurement cycle is aborted.
An ongoing measurement can be aborted with the command ABORT. This command is a query only
and has therefore no *RST value assigned. It is available only when measurement of the maximum
carrier power is selected (see :CONFigure:BURSt:POWer).