FSE READ-Subsystem
1065.6016.12 6.151 E-16
:READ:BURSt:POWer:STATic?
This command increases the static power control level for the measurement by one step, measures
the output power of the base station and reads out the result.
If the command READ:BURSt:POWer:STATic? is repeated after the maximum static power
control level is reached, the measurement sequence is terminated and the result of the maximum
static power control level is read out again. In this case the value FINISHED indicating the status is
read out. Before the status value FINISHED is read out, the value RUNNING is output if the total
result of the limit check is queried via :CALCulate<1|2>:LIMit<1 to 8>:BURSt:POWer?.
Parameter: The result is read out as an ASCII string in the following format:
<Static Power Ctrl>,<Dyn Power Ctrl>,<Rat-Level>,<Act-Level>,
<Delta>,<Status>
<Static Power Ctrl>: current static power control level
<Dyn Power Ctrl>: current dynamic power control level
<Rat-Level>: rated value for the current power control level acc.
to standard in dBm
<Act-Level>: measured power in dBm
<Delta>: difference between the measured power and the power
at the previous static power control level.
<Status>: result of the limit check in character data form:
PASSED no limit exceeded
FAILED limit exceeded
FINISHED measurement sequence terminated
Example: ":READ:BURS:POW:STAT?"
Result: 1,0,41,42.5,1.6,PASSED
Features: *RST value: --
SCPI: device-specific
Mode: BTS
The command ABORt terminates an ongoing measurement and resets the static and dynamic power
control level to 0.
This command is only a query and therefore has no *RST value assigned. It is available only if
measurement of the maximum carrier power is selected. (see :CONFigure:BURSt:POWer).