READ-Subsystem FSE
1065.6016.12 6.152 E-16
:READ:BURSt:POWer:DYNamic?
This command increases the dynamic power control level for the measurement by one step,
measures the output power of the base station and reads out the result.
Once the maximum dynamic power control level is reached the command is accepted only after the
static power control level is increased by one step.
Note that the command is no longer accepted after the measurement sequence is terminated which
implies that the static power control level was read out again with READ:BURSt:POWer:STATic?
after the maximum value was reached and marked with the FINISHED status.
Parameter: The result is read out as an ASCII string in the following format:
<Static Power Ctrl>,<Dyn Power Ctrl>,<Rat-Level>,<Act-Level>,
<Delta>,<Status>
<Static Power Ctrl>: current static power control level
<Dyn Power Ctrl>: current dynamic power control level
<Rat-Level>: rated value for the current power control level acc.
to standard in dBm
<Act-Level>: measured power in dBm
<Delta>: difference between the measured power and the power
at the previous dynamic power control level.
<Status>: result of the limit check in character data form:
PASSED no limit exceeded
FAILED limit exceeded
Example: ":READ:BURS:POW:DYN?"
Result: 1,3,35,32.5,5.6,FAILED
Features: *RST value: --
SCPI: device-specific
Mode: BTS
The command ABORt terminates an ongoing measurement and resets the static and dynamic power
control level to 0.
This command is only a query and therefore has no *RST value assigned. It is available only if
measurement of the maximum carrier power is selected. (see :CONFigure:BURSt:POWer).