EasyManua.ls Logo

Rohde & Schwarz FSEA20 - Page 210

Rohde & Schwarz FSEA20
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FSE READ-Subsystem
1065.6016.12 6.153 E-16
:READ:BURSt:POWer:LEVel?
This command increases the power control level for the measurement by one step, measures the
output power of the mobile and reads out the result.
Note that the command is no longer accepted after the measurement sequence is terminated which
implies that the power control level was read out again with READ:BURSt:POWer:LEVel? after the
maximum value was reached and marked with the FINISHED status.
Parameter: The result is read out as an ASCII string in the following format:
<0>, <Power Ctrl Level>,<Rat-Level>,<Act-Level>, <Delta>,<Status>
<0>: always 0
<Power Ctrl Level>: current power control level
<Rat-Level>: rated value for the current power control level acc.
to standard in dBm
<Act-Level>: measured power in dBm
<Delta>: difference between the measured power and the power
at the previous power control level.
<Status>: result of the limit check in character data form:
PASSED no limit exceeded
FAILED limit exceeded
Example: ":READ:BURS:POW:LEV?"
Result: 0,3,35,32.5,5.6,FAILED
Features: *RST value: --
SCPI: device-specific
Mode: MS
The command ABORt terminates an ongoing measurement and resets the power control level to 0.
This command is only a query and therefore has no *RST value assigned. It is available only if
measurement of the maximum carrier power is selected. (see :CONFigure:BURSt:POWer).
:READ:BURSt:REFerence[:IMMediate]?
This command starts the premeasurement and as a result provides the measured level in dBm.
Example: ":READ:BURS:REF?"
Feature: *RST value: --
SCPI: device-specific
Mode: MS, BTS
This is a query command only and therefore has no *RST value.

Table of Contents

Related product manuals