EasyManua.ls Logo

Rohde & Schwarz FSEA20 - Page 214

Rohde & Schwarz FSEA20
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FSE READ-Subsystem
1065.6016.12 6.157 E-16
:READ:BURSt:MACCuracy:PERCentile:STATus?
This command starts the measurement of the modulation accuracy of the base station or mobile and
reads out the status of the 95% percentile measurement taken over the selected number of bursts.
0: failed, 1: passed
Example: ":READ:BURS:MACC:PERC:STAT?"
Features: *RST value: --
SCPI: device-specific
Modes:BTS, MS
When the measurement is started the instrument automatically assumes the SINGLE mode.
An ongoing measurement can be aborted via the command ABORt. Further results of the modulation
accuracy measurement can be then queried without restart of the measurement via the
:FETCh:BURSt-subsystem.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see :CONFigure:BURSt:MACCuracy).
:READ:BURSt:MACCuracy:PERCentile:AVERage?
This command starts the measurement of the modulation accuracy of the base station or mobile and
reads out the average of the 95% percentile measurement taken over the selected number of bursts.
Example: ":READ:BURS:MACC:PERC:AVER?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
When the measurement is started the instrument automatically assumes the SINGLE mode.
An ongoing measurement can be aborted via the command ABORt. Further results of the modulation
accuracy measurement can be then queried without restart of the measurement via the
:FETCh:BURSt-subsystem.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see :CONFigure:BURSt:MACCuracy).
:READ:BURSt:MACCuracy:PERCentile:MAXimum?
This command starts the measurement of the modulation accuracy of the base station or mobile and
reads out the maximum of the 95% percentile measurement for the selected number of bursts.
Example: ":READ:BURS:MACC:PERC:MAX?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
When the measurement is started the instrument automatically assumes the SINGLE mode.
An ongoing measurement can be aborted via the command ABORt. Further results of the modulation
accuracy measurement can be then queried without restart of the measurement via the
:FETCh:BURSt-subsystem.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see :CONFigure:BURSt:MACCuracy).

Table of Contents

Related product manuals