READ-Subsystem FSE
1065.6016.12 6.158 E-16
:READ:BURSt:MACCuracy:FREQuency:STATus?
This command starts the measurement of the modulation accuracy of the base station or mobile and
reads out the status of the frequency error measurement taken over the selected number of bursts.
0: failed, 1: passed
Example: ":READ:BURS:MACC:FREQ:STAT?"
Features: *RST value: --
SCPI: device-specific
Modes:BTS, MS
When the measurement is started the instrument automatically assumes the SINGLE mode.
An ongoing measurement can be aborted via the command ABORt. Further results of the modulation
accuracy measurement can be then queried without restart of the measurement via the
:FETCh:BURSt-subsystem.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see :CONFigure:BURSt:MACCuracy).
:READ:BURSt:MACCuracy:FREQuency:AVERage?
This command starts the measurement of the modulation accuracy of the base station or mobile and
reads out the average of the frequency error measurement taken over the selected number of bursts.
Example: ":READ:BURS:MACC:FREQ:AVER?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
When the measurement is started the instrument automatically assumes the SINGLE mode.
An ongoing measurement can be aborted via the command ABORt. Further results of the modulation
accuracy measurement can be then queried without restart of the measurement via the
:FETCh:BURSt-subsystem.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see :CONFigure:BURSt:MACCuracy).
:READ:BURSt:MACCuracy:FREQuency:MAXimum?
This command starts the measurement of the modulation accuracy of the base station or mobile and
reads out the maximum of the frequency error measurement for the selected number of bursts.
Example: ":READ:BURS:MACC:FREQ:MAX?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
When the measurement is started the instrument automatically assumes the SINGLE mode.
An ongoing measurement can be aborted via the command ABORt. Further results of the modulation
accuracy measurement can be then queried without restart of the measurement via the
:FETCh:BURSt-subsystem.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see :CONFigure:BURSt:MACCuracy).