SENSe Subsystem FSE
1065.6016.12 6.172 E-16
:[SENSe<1|2>:]CORRection[:STATe] ON | OFF
This command activates/deactivates normalization of the measurement results.
Example: ":CORR ON "
Features: *RST value: OFF
SCPI: conforming
Mode: A
This command is only valid in conjunction with option Tracking Generator.
:[SENSe<1|2>:]CORRection:METHod TRANsmission | REFLexion
This command selects the kind of measurement with active tracking generator
(transmission/reflexion).
Example: ":CORR:METH TRAN "
Features: *RST value: TRANsmission
SCPI: device specific
Mode: A
This command is only valid in conjunction with option Tracking Generator.
:[SENSe<1|2>:]CORRection:COLLect[:ACQuire] THRough | OPEN
This command selects the kind of measurement for the reference values of the normalization
(response calibration).
Example: ":CORR:COLL THR"
Features: *RST value:
SCPI: conforming
Mode: A
THRough "TRANsmission" mode: calibration with direct connection between tracking
generator and device input.
"REFLexion" mode: calibration with short circuit at the input
OPEN only valid in "REFLexion" mode:calibration with open input
This command is an event which is why it is not assigned an *RST value an a query. It is only valid in
conjunction with option Tracking Generator.
:[SENSe<1|2>:]CORRection:RECall
This command restores the instrument setting that was valid for the measurement of the reference
data.
Example: ":CORR:REC"
Features: *RST value: -
SCPI: conforming
Mode: A
This command is an event which is why it is not assigned an *RST value and a query. It is only valid
in conjunction with option Tracking Generator.