FSE SENSe Subsystem
1065.6016.12 6.177 E-16
:[SENSe<1|2>:]CORRection:TSET[:STATe] ON | OFF
This command switches the selected transducer set on or off.
Example: ":CORR:TSET ON"
Features: *RST value: OFF
SCPI: device-specific
Mode: A
Prior to this command, the command SENS:CORR:TSET:SEL must be sent.
:[SENSe<1|2>:]CORRection:TSET:DELete
This command deletes the selected transducer set.
Example: ":CORR:TSET:DEL"
Features: *RST value: -
SCPI: device-specific
Mode: A
This command is an event and thus has no *RST value assigned.
Prior to this command, the command SENS:CORR:TSET:SEL must be sent.
:[SENSe<1|2>:]CORRection:LOSS:INPut[:MAGNitude] <numeric_value>
This command announces to the instrument a possibly needed external attenuation of the input
signal, so that it is taken into account later when the level is set.
Parameter: <numeric_value>::= value of external attenuation in dB.
Example: ":CORR:LOSS:INP 30DB "
Features: *RST value: 20dB
SCPI: device-specific
Modes: BTS, MS
The external attenuation must be selected such that the input power of the analyzer does not exceed
27 dBm.
:[SENSe<1|2>:]CORRection:RXGain:INPut[:MAGNitude] <numeric_value>
This command announces to the instrument a possibly needed preamplification in the RX-band (RX
BAND GAIN), so that it is taken into account later when the measured values are read out.
Parameter: <numeric_value>::= value of the amplification in dB.
Example: ":CORR:RXG:INP 30DB "
Features: *RST value: 0 dB
SCPI: device-specific
Modes: BTS, MS