Loading...
SENSe Subsystem FSE
1065.6016.12 6.178 E-16
:[SENSe<1|2>:]CORRection:CVL:CATalog?
This command polls the names of all conversion-loss tables stored on the harddisk.
Syntax of output format:
<Sum of file lengths of all subsequent files>,<free memory on hard disk>,
<1st file name>,,<1st file length>,<2nd file name>,,<2nd file length>,....,<nth file name>,
<nth file length>
Example: ":CORR:CVL:CAT?"
Feature: *RST value: -
SCPI: device-specific
Mode: A
This command is only valid in conjunction with option External Mixer Output, FSE-B21.
:[SENSe<1|2>:]CORRection:CVL:SELect <file_name>
This command selects the Conversion Loss Table designated with <file_name>. If <file_name> is
not available, a new Conversion Loss Table will be created.
Parameter: <file_name>::= Name of Conversion Loss Table as string data with
a maximum of 8 characters
Example: ":CORR:CVL:SEL LOSS_TAB’"
Features: *RST value: -
SCPI: device-specific
Mode: A
This command must be sent prior to the subsequent commands used to change/activate the
Conversion Loss files. It is only valid in conjunction with option External Mixer Output, FSE-B21.
:[SENSe<1|2>:]CORRection:CVL:MIXer <string>
This command defines the type designation of the mixer in the Conversion Loss Table.
Parameter: <string>::= Type designation of mixer with a maximum of 16 characters
Example: ":CORR:CVL:MIX FSE_Z60’"
Features: *RST value: -
SCPI: device-specific
Mode: A
Command SENS:CORR:CVL:SEL must be sent prior to this command. This command is only valid
in conjunction with option External Mixer Output, FSE-B21.