FSE SENSe Subsystem
1065.6016.12 6.179 E-16
:[SENSe<1|2>:]CORRection:CVL:SNUMber <string>
This command defines the serial number of the mixer in the Conversion Loss Table.
Parameter: <string>::= Serial number of mixer with a maximum of 16 characters
Example: ":CORR:CVL:SNUM 123.4567’"
Features: *RST value: -
SCPI: device-specific
Mode: A
Command SENS:CORR:CVL:SEL must be sent prior to this command. This command is only valid
in conjunction with option External Mixer Output, FSE-B21.
:[SENSe<1|2>:]CORRection:CVL:BAND A|Q|U|V|E|W|F|D|G|Y|J
This command defines the waveguide band in the Conversion Loss Table.
Example: ":CORR:CVL:BAND E"
Features: *RST value: -
SCPI: device-specific
Mode: A
Command SENS:CORR:CVL:SEL must be sent prior to this command. This command is only valid
in conjunction with option External Mixer Output, FSE-B21.
:[SENSe<1|2>:]CORRection:CVL:TYPE ODD | EVEN | EODD
This command defines the type of harmonic in the Conversion Loss Table.
Example: ":CORR:CVL:TYPE EODD"
Features: *RST value: -
SCPI: device-specific
Mode: A
Command SENS:CORR:CVL:SEL must be sent prior to this command. This command is only valid
in conjunction with option External Mixer Output, FSE-B21.
:[SENSe<1|2>:]CORRection:CVL:PORTs 2 | 3
This command defines the type of mixer in the Conversion Loss Table.
Example: ":CORR:CVL:PORT 3"
Features: *RST value: -
SCPI: device-specific
Mode: A
Command SENS:CORR:CVL:SEL must be sent prior to this command. This command is only valid
in conjunction with option External Mixer Output, FSE-B21.