SENSe Subsystem FSE
1065.6016.12 6.194 E-16
:[SENSe<1|2>:]FREQuency:CENTer:STEP:LINK SPAN | RBW | OFF
This command couples the step width of the center frequency to span (span >0) or to the resolution
bandwidth (span = 0) or cancels the couplings.
Example: ":FREQ:CENT:STEP:LINK SPAN"
Features: *RST value: SPAN
SCPI: device-specific
Modes: A, BTS, MS
:[SENSe<1|2>:]FREQuency:CENTer:STEP:LINK:FACTor 1 to 100 PCT
This command couples the step width of the center frequency with a factor to the span (span >0) or
to the resolution bandwidth (span = 0).
Example: ":FREQ:CENT:STEP:LINK:FACT 20PCT"
Features: *RST value: - (AUTO 0.1 × SPAN is switched on)
SCPI: device-specific
Modes: A, BTS, MS
:[SENSe<1|2>:]FREQuency:SPAN 0 GHz to f
max
This command defines the frequency span of the analyzer.
Example: ":FREQ:SPAN 10MHz"
Features: *RST value: f
max
SCPI: conforming
Mode: A
The automatic coupling of the parameters is set to CENTER FIXED.
:[SENSe<1|2>:]FREQuency:SPAN:FULL
This command sets the maximum frequency span of the analyzer.
Example: ":FREQ:SPAN:FULL"
Features: *RST value: -
SCPI: conforming
Mode: A
This command is an event which is why it is not assigned an *RST value and has no query.
:[SENSe<1|2>:]FREQuency:SPAN:LINK CENTer | START | STOP
This command defines the coupling for frequency-span changes.
Example: ":FREQ:SPAN:LINK STOP"
Features: *RST value: CENTer
SCPI: conforming
Mode: A