FSE SENSe Subsystem
1065.6016.12 6.195 E-16
:[SENSe<1|2>:]FREQuency:STARt 0 GHz to f
max
This command defines the start frequency of the analyzer.
Example: ":FREQ:STAR 20MHz"
Features: *RST value: 0
SCPI: conforming
Mode: A
The automatic coupling of the parameters is set to STOP FIXED.
:[SENSe<1|2>:]FREQuency:STARt:LINK CENTer | STOP | SPAN
This command defines the coupling for start-frequency changes.
Example: ":FREQ:STAR:LINK SPAN"
Features: *RST value: STOP
SCPI: device-specific
Mode: A
:[SENSe<1|2>:]FREQuency:STOP 0 GHz to f
max
This command defines the stop frequency of the scan in receiver mode or the stop frequency of the
analyzer.
Example: ":FREQ:STOP 2000MHz"
Features: *RST value: f
max
SCPI: conforming
Mode: A
The automatic coupling of the parameters is set to STARt FIXED.
:[SENSe<1|2>:]FREQuency:STOP:LINK CENTer | STARt | SPAN
This command defines the coupling for stop-frequency changes.
Example: ":FREQ:STOP:LINK SPAN"
Features: *RST value: STARt
SCPI: device-specific
Mode: A