SENSe Subsystem FSE
1065.6016.12 6.196 E-16
:[SENSe<1|2>:]FREQuency:MODE CW | FIXed | SWEep
This command switches between frequency (SWEep) and time (CW | FIXed) domain in the analyzer
mode.
Example: ":FREQ:MODE SWE"
Features: *RST value: CW
SCPI: conforming
Mode: A
For CW and FIXed, the frequency setting is via command
FREQuency:CENTer. In the SWEep
mode, the setting is via commands
FREQuency:STARt, STOP, CENTer and SPAN.
:[SENSe<1|2>:]FREQuency:OFFSet <numeric_value>
This command defines the frequency offset of the instrument.
Example: ":FREQ:OFFS 1GHZ"
Features: *RST value: 0 Hz
SCPI: conforming
Modes: A, VA