SENSe Subsystem FSE
1065.6016.12 6.208 E-16
:[SENSe<1|2>:]SWEep:COUNt 0 to 32767
This command defines the number of sweeps started with single sweep.
Example: ":SWE:COUN 64"
Features: *RST value: 0
SCPI: conforming
Modes: A, VA-D
This parameter defines the number of sweeps or the number of averaging procedures. In the
average mode, the value 0 defines a running averaging of measurement data over 10 sweeps.
:[SENSe<1|2>:]SWEep:EGATe ON | OFF
This command controls the sweep with the external gate signal.
Example: ":SWE:EGAT ON"
Features: *RST value: OFF
SCPI: device-specific
Mode: A
:[SENSe<1|2>:]SWEep:EGATe:LEVel -5V to +5V
This command determines the threshold for the external gate signal.
Example: ":SWE:EGAT:LEV 3V"
Features: *RST value: 2V
SCPI: device-specific
Mode: A
If SWEep:TIME is directly programmed, automatic coupling is switched off.
:[SENSe<1|2>:]SWEep:EGATe:TYPE LEVel | EDGE
This command sets the type of triggering (level or edge) by the external gate signal.
Example: ":SWE:EGAT:TYPE EDGE"
Features: *RST value: EDGE
SCPI: device-specific
Mode: A
Parameter EGATe:LENGth is not used for level triggering.
:[SENSe<1|2>:]SWEep:EGATe:POLarity POSitive | NEGative
This command determines the polarity of the external gate signal.
Example: ":SWE:EGAT:POL POS"
Features: *RST value: POSitive
SCPI: device-specific
Mode: A