FSE SENSe Subsystem
1065.6016.12 6.209 E-16
:[SENSe<1|2>:]SWEep:EGATe:HOLDoff 0 to 100s
This command defines the delay time between the external gate signal and the continuation of the
sweep.
Example: ":SWE:EGAT:HOLD 100us"
Features: *RST value: 0s
SCPI: device-specific
Mode: A
The values for the delay time can be set in steps 1, 2, 3 and 5.
:[SENSe<1|2>:]SWEep:EGATe:LENGth 0 to 100s
In case of edge triggering, this command determines the time interval in which the instrument
sweeps.
Example: ":SWE:EGAT:LENG 10ms"
Features: *RST value: 0s
SCPI: device-specific
Modes: A
The values for the delay time can be set in steps 1, 2, 3 and 5.
:[SENSe<1|2>:]SWEep:EGATe:SOURce EXTernal | RFPower
This command toggles between external gate signal and RF power signal.
Example: ":SWE:EGAT:SOUR RFP"
Features: *RST value: EXTernal
SCPI: device-specific
Mode: A
:[SENSe<1|2>:]SWEep:GAP ON | OFF
This command controls the operating mode GAP SWEEP.
Example: ":SWE:GAP ON"
Features: *RST value: OFF
SCPI: device-specific
Mode: A-Z
Operating mode GAP SWEEP for measurements in the time domain controls the display of
measured values. Measured values can be blanked for a defined time range (GAP).
:[SENSe<1|2>:]SWEep:GAP:PRETrigger 0 to 100s
This command defines the evaluation time for measured values before the pretrigger time
(resolution: 50 ns).
Example: ":SWE:GAP:PRET 100us"
Features: *RST value: 0s
SCPI: device-specific
Mode: A