SENSe Subsystem FSE
1065.6016.12 6.210 E-16
:[SENSe<1|2>:]SWEep:GAP:TRGTogap 0 to 100s
This command defines the time between the pretrigger time and the beginning of the gap (trigger-to-
gap time) (resolution: 50 ns).
Example: ":SWE:GAP:TRGT 50us"
Features: *RST value: 0s
SCPI: device-specific
Mode: A
:[SENSe<1|2>:]SWEep:GAP:LENGth 0 to 100s
This command defines the gap length.
Example: ":SWE:GAP:LENG 400us"
Features: *RST value: 0s
SCPI: device-specific
Mode: A
The gap length can be programmed from 0 to 100s in steps of 1, 2, 3 and 5.
:[SENSe<1|2>:]SWEep:SPACing LINear | LOGarithmic
This command toggle between linear and logarithmic sweep.
Example: ":SWE:SPAC LOG"
Features: *RST value: LIN
SCPI: conforming
Modes: A,
The frequency axis is set to linear or logarithmic scaling accordingly.