CALCulate Subsystem FSE
1065.6016.12 6.10 E-16
:CALCulate<1|2>:DELTamarker<1 to 4>:MODE ABSolute | RELative
This command switches over between relative and absolute input of frequency of the delta marker.
Example: ":CALC:DELT:MODE ABS"
Features: *RST value: REL
SCPI: device-specific
Modes: A, VA, BTS, MS
In the RELative mode, the frequency of the delta marker is programmed relative to the reference
marker. In the ABSolute mode, the frequency is defined by the absolute values.
:CALCulate<1|2>:DELTamarker<1 to 4>:AOFF
This command switches off all active delta markers.
Example: ":CALC:DELT:AOFF"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
:CALCulate<1|2>:DELTamarker<1 to 4>:TRACe 1 to 4
This command assigns the selected delta marker to the indicated measuring curve.
Example: ":CALC:DELT3:TRAC 2"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
:CALCulate<1|2>:DELTamarker<1 to 4>:X 0 to MAX (frequency | sweep time | symbols)
This command positions the selected delta marker to the indicated frequency (span > 0) or time
(span = 0). The query always returns the absolute value of frequency or time.
Example: ":CALC:DELT:X 10.7MHz"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
The SYM unit is only valid in Vector Signal Analysis mode.
:CALCulate<1|2>:DELTamarker<1 to 4>:X:RELative?
This command queries the frequency (span > 0) or time (span = 0) of the selected delta marker
relative to the reference marker.
Example: ":CALC:DELT:X:REL?"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS