FSE CALCulate Subsystem
1065.6016.12 6.11 E-16
:CALCulate<1|2>:DELTamarker<1 to 4>:Y?
This command queries the value of the selected marker.
Example: ":CALC:DELT:Y?"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
In complex presentations (vector signal analysis - polar diagrams), the real and the imaginary
component as well as magnitude and phase are output separated by a comma.
:CALCulate<1|2>:DELTamarker<1 to 4>:MAXimum[:PEAK]
This command positions the delta marker to the current maximum value in the trace memory.
Example: ":CALC:DELT:MAX"
Features: *RST value: -
SCPI: device-specific
Modes: A, VA, BTS, MS
This command is an event which is why it is not assigned an *RST value and has no query.
:CALCulate<1|2>:DELTamarker<1 to 4>:MAXimum:APEak
This command positions the delta marker to the maximum absolute value of the trace.
Example: ":CALC:DELT:MAX:APE"
Features: *RST value: -
SCPI: device-specific
Modes: VA
This command is an event which is why it is not assigned an *RST value and has no query.
:CALCulate<1|2>:DELTamarker<1 to 4>:MAXimum:NEXT
This command positions the delta marker to the next smaller maximum value in the trace memory.
Example: ":CALC:DELT:MAX:NEXT"
Features: *RST value: -
SCPI: device-specific
Modes: A, BTS, MS
This command is an event which is why it is not assigned an *RST value and has no query.
:CALCulate<1|2>:DELTamarker<1 to 4>:MAXimum:RIGHt
This command positions the delta marker to the next smaller maximum value to the right of the
current value (i.e., in ascending X direction) in the trace memory.
Example: ":CALC:DELT:MAX:RIGH"
Features: *RST value: -
SCPI: device-specific
Modes: A, BTS, MS
This command is an event which is why it is not assigned an *RST value and has no query.