CALCulate Subsystem FSE
1065.6016.12 6.14 E-16
:CALCulate<1|2>:DELTamarker<1 to 4>:FUNCtion:PNOise:RESult?
This command queries the result of the phase noise measurement.
Example: ":CALC:DELT:FUNC:PNO:RES?"
Features: *RST value: -
SCPI: device-specific
Modes: A, BTS, MS
This command is only a query which is why it is not assigned an *RST value.
:CALCulate<1|2>:DELTamarker<1 to 4>:STEP[:INCRement] <numeric_value>
This command defines the delta marker step width.
Example: ":CALC:DELT:STEP 10kHz" (frequency domain)
":CALC:DELT:STEP 5ms" (time domain)
Features: *RST value: - (STEP is set to AUTO)
SCPI: device-specific
Mode: A
:CALCulate<1|2>:DELTamarker<1 to 4>:STEP:AUTO ON | OFF
This command switches the automatic adaptation of the marker step width on or off.
Example: ":CALC:DELT:STEP:AUTO OFF"
Features: *RST value: ON
SCPI: device-specific
Mode: A
With AUTO ON, the step width is 10% of the span.