FSE CALCulate Subsystem
1065.6016.12 6.23 E-16
:CALCulate<1|2>:LIMit<1 to 8>:CONTrol:DOMain FREQuency | TIME
This command defines the X-axis in the frequency or time domain.
Example: ":CALC:LIM:CONT:DOM TIME"
Features: *RST value: FREQuency
SCPI: device-specific
Modes: A, VA, BTS, MS
:CALCulate<1|2>:LIMit<1 to 8>:CONTrol:OFFSet <numeric_value>
This command defines an offset for the X-axis value of the selected relative limit line in the frequency
or time domain.
Example: ":CALC:LIM:CONT:OFFS 100us"
Features: *RST value: 0
SCPI: device-specific
Modes: A, VA
:CALCulate<1|2>:LIMit<1 to 8>:CONTrol:MODE RELative | ABSolute
This command selects the relative or absolute scaling for the X-axis of the selected limit line.
Example: ":CALC:LIM:CONT:MODE REL"
Features: *RST value: ABSolute
SCPI: device-specific
Modes: A, VA, BTS, MS
Upon selection of RELative, the unit is switched to DB.
:CALCulate<1|2>:LIMit<1 to 8>:CONTrol:UNIT[:TIME] S | SYM
This command defines the unit of the x-axis scaling of limit lines.
Example: ":CALC:LIM:CONT:UNIT SYM"
Features: *RST value: S
SCPI: device-specific
Mode: VA
:CALCulate<1|2>:LIMit<1 to 8>:CONTrol:SHIFt <numeric_value>
This command shifts a limit line which has been specified for relative frequencies or times (X-axis).
Example: ":CALC:LIM:CONT:SHIF 50kHz"
Features: *RST value: --
SCPI: device-specific
Modes: A, VA, BTS, MS
This command is an event which is why it is not assigned an *RST value and has no query.