CALCulate Subsystem FSE
1065.6016.12 6.22 E-16
:CALCulate<1|2>:LIMit<1 to 8>:STATe ON | OFF
This command switches the limit check for the selected limit line on or off. With limit check switched
off, the limit line is disabled.
Example: ":CALC:LIM:STAT ON"
Features: *RST value: OFF
SCPI: conforming
Modes: A, VA, BTS, MS
The result of the limit check can be queried with CALCulate:LIMit:FAIL?.
:CALCulate<1|2>:LIMit<1 to 8>:UNIT DBM | DBPW | DBPT | WATT | DBUV | DBMV | VOLT |DBUA
| AMPere | DB | DBUV_MHZ | DBMV_MHZ | DBUA_MHZ |
DBUV_M | DBUA_M | DBUV_MMHZ | DBUA_MMHZ |
UNITLESS|
This command defines the unit of the selected limit line.
Example: ":CALC:LIM:UNIT DBUV"
Features: *RST value: DBM
SCPI: device-specific
Modes: A, VA, BTS, MS
DBUV_MHZ and DBUA_MHZ denote the units DBUV/MHZ or DBUA/MHZ.
Upon selection of the unit DB the limit line is automatically switched to the relative mode. For units
different from DB the limit line is automatically switched to the absolute mode.
The units DEG, RAD, S, HZ are available in the vector analysis mode only.
:CALCulate<1|2>:LIMit:CATalog?
This command reads out the names of all limit lines stored on the harddisk.
Syntax of output format:
<Sum of file lengths of all subsequent files>,<free memory on hard disk>,
<1st file name>,,<1st file length>,<2nd file name>,,<2nd file length>,....,<nth file name>,
<nth file length>
Example: ":CALC:LIM:CAT?"
Feature: *RST value: -
SCPI: device-specific
Mode: A, VA, BTS, MS
:CALCulate<1|2>:LIMit<1 to 8>:CONTrol[:DATA] <numeric_value>,<numeric_value>..
This command defines the X-axis values (frequencies or times) of the upper or lower limit lines.
Example: ":CALC:LIM:CONT 1MHz,30MHz,300MHz,1GHz"
Features: *RST value: - (LIMit:STATe is set to OFF)
SCPI: conforming
Modes: A, VA, BTS, MS
The number of values for the CONTrol axis and the corresponding UPPer- and/or LOWer limit lines
have to be identical. Available units are HZ | S | SYM, SYM only for vector signal analyzer mode.