FSE CALCulate Subsystem
1065.6016.12 6.27 E-16
:CALCulate<1|2>:LIMit<1 to 8>:LOWer:SPACing LINear | LOGarithmic
This command makes a selection between linear and logarithmic interpolation for the lower limit line.
Example: ":CALC:LIM:LOW:SPAC LIN"
Features: *RST value: LIN
SCPI: device-specific
Modes: A, VA, BTS, MS
:CALCulate<1|2>:LIMit<1 to 8>:FAIL?
This command queries the result of the limit check.
Example: ":CALC:LIM:FAIL?"
Features: *RST value: -
SCPI: conforming
Modes: A, VA, BTS, MS
The result of the limit check responds with 0 in case of PASS and with 1 in case of FAIL.
For measurements spectrum due to modulation and spectrum due to transients (options FSE-K10
and FSE-K11), the result of the limit check is queried with this command in frequency sweep mode.
:CALCulate<1|2>:LIMit<1 to 8>:CLEar[:IMMediate]
This command deletes the result of the current limit check.
Example: ":CALC:LIM:CLE"
Features: *RST value: -
SCPI: conforming
Modes: A, VA, BTS, MS
This command is an event which is why it is not assigned an *RST value.
:CALCulate<1|2>:LIMit<1 to 8>:COMMent <string>
This command defines a comment for the limit line selected.
Example: ":CALC:LIM:COMM ’Upper limit for spectrum’"
Features: *RST value: blank comment
SCPI: device-specific
Modes: A, VA, BTS, MS