EasyManua.ls Logo

Rohde & Schwarz FSEA20 - Page 85

Rohde & Schwarz FSEA20
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CALCulate Subsystem FSE
1065.6016.12 6.28 E-16
:CALCulate<1|2>:LIMit<1 to 8>:COPY 1 to 8 | <name>
This command copies one limit line onto another one.
Parameter: 1 to 8 ::= number of the new limit line or, alternatively:
<name> ::= name of the new limit line given as a string
Example: ":CALC:LIM1:COPY 2"
":CALC:LIM1:COPY GSM2’"
Features: *RST value: --
SCPI: device-specific
Modes: A, VA, BTS, MS
The name of the limit line may contain a maximum of 8 characters. This command is an "event"
which is why it is not assigned an *RST value and has no query.
:CALCulate<1|2>:LIMit<1 to 8>:NAME <name of limit line>
This command assigns a name to a limit line numbered 1 to 8. If a limit line of the given name
doesnt exist previously, a limit line with this name is created. The values of a previous limit line with
the selected line number are kept and the current unit is used. If no limit line with the selected line
number was defined yet, the correct values for the x and y axis have to be entered before the new
limit line will be saved (using commands CALCulate:LIMit:CONTrol:DATA and
CALCulate:LIMit:LOWer|UPPer:DATA).
Example: ":CALC:LIM1:NAME GSM1’"
Features: *RST value: REM1 to REM8 for lines 1 to 8
SCPI: device-specific
Modes: A, VA, BTS, MS
The name of the limit line may contain a maximum of 8 characters.
:CALCulate<1|2>:LIMit<1 to 8>:DELete
This command deletes the limit line selected.
Examples: ":CALC:LIM1:DEL"
Features: *RST value: --
SCPI: device-specific
Modes: A, VA, BTS, MS
This command is an "event" which is why it is not assigned an *RST value and has no query.