FSE CALCulate Subsystem
1065.6016.12 6.29 E-16
:CALCulate<1|2>:LIMit<1 to 8>:BURSt:PTEMplate?
This command queries the result of the limit check for a power vs. time measurement.
Parameter: The result is displayed in character data form. Possible values are:
PASSED limit not exceeded
FAILED limit exceeded
RUNNING measurement not completed
Examples: ":CALC:LIM:BURS:PTEM?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
This command is a query and therefore not assigned a *RST value.
If no measurement has been carried out yet, a query error is triggered off. The numeric suffixes
<1|2> or <1 to 8> are not significant for this command.
:CALCulate<1|2>:LIMit<1 to 8>:BURSt:POWer?
This command queries the total result of the carrier power measurement.
Parameter: The result is displayed in character data form. Possible values are:
PASSED limit not exceeded
FAILED limit exceeded
ABORTED measurement aborted
RUNNING measurement not completed
Examples: ":CALC:LIM:BURS:POW?"
Result: PASSED
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
This command is a query and therefore not assigned a *RST value.
If the command is triggered off before the carrier power measurement was started for the first time, a
query error results. The numeric suffixes <1|2> or <1 to 8> are not significant for this command.
:CALCulate<1|2>:LIMit<1 to 8>:BURSt:PFERror?
This command queries the total result of the phase/frequency measurement.
Parameter: Result 1 limit not exceeded
0 limit exceeded
Example: ":CALC:LIM:BURS:PFER?"
Result:1
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
This command is a query and therefore not assigned a *RST value.
If the command is triggered off before the phase/frequency measurement was started for the first
time, a query error results. The numeric suffixes <1|2> or <1 to 8> are not significant for this
command.