FSE CALCulate Subsystem
1065.6016.12 6.33 E-16
:CALCulate<1|2>:LIMit<1 to 8>:SPURious:FAILs? TXBand | OTXBand | RXBand | IDLeband
This command queries the number of limit violations of the spurious emissions measurement.
Examples: ":CALC:LIM:SPUR:FAIL? OTXB"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
TXBand TX-band
OTXBand Not TX-band
RXBand RX-band (option FSE-K11 only)
IDLeband IDLeband (option FSE-K10 only)
This command is a query and therefore not assigned a *RST value. The numeric suffixes <1|2> or
<1 to 8> are not significant for this command.
:CALCulate<1|2>:LIMit<1 to 8>:MARGin 0 to 100DB
This command sets/changes the value of the margin (safe difference to the actual limit) for the limit
check.
Examples: ":CALC:LIM:MARG 6DB"
Features: *RST value: 3DB
SCPI: device-specific
Modes: BTS, MS
The numeric suffixes <1|2> or <1 to 8> are not significant for this command.
:CALCulate<1|2>:LIMit<1 to 8>:ACPower[:STATe] ON | OFF
This command switches on and off the limit check for adjacent channel power measurements. The
commands CALC:LIM:ACP:ACH:STAT or CALC:LIM:ACP:ALT:STAT must be used in addition to
specify whether the limit check is to be performed for the upper/lower adjacent channel or for the
alternate adjacent channels.
Examples: ":CALC:LIM:ACP ON"
Features: *RST value: OFF
SCPI: device-specific
Modes: A, VA
The numeric suffixes <1|2> or <1 to 8> are not significant for this command.
:CALCulate<1|2>:LIMit<1 to 8>:ACPower:ACHannel 0 to 100 dB, 0 to 100 dB
This command defines the limit for the upper/lower adjacent channel for adjacent channel power
measurements.
Parameter: The first (second) numeric value is the limit for the upper (lower) adjacent
channel.
Examples: ":CALC:LIM:ACP:ACH 30DB, 30DB"
Features: *RST value: 0 dB
SCPI: device-specific
Modes: A, VA
The numeric suffixes <1|2> or <1 to 8> are not significant for this command.