CALCulate Subsystem FSE
1065.6016.12 6.32 E-16
:CALCulate<1|2>:LIMit<1 to 8>:SPECtrum:SWITching?
This command queries the total result of the spectrum due to switching transients measurements for
list mode. For frequency mode, the limit violations are queried with command
CALCulate:LIMit:FAIL?
Parameter: The result is displayed in character data form. Possible values are:
PASSED limit not exceeded
FAILED limit exceeded
ABORTED measurement aborted
RUNNING measurement not completed
Examples: ":CALC:LIM:SPEC:SWIT?"
Result: PASSED
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
This command is a query and therefore not assigned a *RST value. The numeric suffixes <1|2> or
<1 to 8> are not significant for this command.
:CALCulate<1|2>:LIMit<1 to 8>:SPECtrum:SWITching:FAILs?
This command queries the number of limit violations of the spectrum due to switching transient
measurement for list mode.The number of limit violations is the total of all violations above and below
the carrier. For frequency mode, the limit violations are queried with command
CALCulate:LIMit:FAIL?.
Examples: ":CALC:LIM:SPEC:SWIT:FAIL?"
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
This command is a query and therefore not assigned a *RST value. The numeric suffixes <1|2> or
<1 to 8> are not significant for this command.
:CALCulate<1|2>:LIMit<1 to 8>:SPURious? TXBand | OTXBand | RXBand | IDLeband
This command queries the total result of the spurious emissions measurement.
Parameter: The result is displayed in character data form. Possible values are:
PASSED limit not exceeded
FAILED limit exceeded
ABORTED measurement aborted
RUNNING measurement not completed
Examples: ":CALC:LIM:SPUR? OTXB"
Result:PASSED
Features: *RST value: --
SCPI: device-specific
Modes: BTS, MS
TXBand TX-band
OTXBand Not TX-band
RXBand RX-band (option FSE-K11 only)
IDLeband IDLeband (option FSE-K10 only)
This command is a query and therefore not assigned a *RST value. The numeric suffixes <1|2> or
<1 to 8> are not significant for this command.