CALCulate Subsystem FSE
1065.6016.12 6.40 E-16
:CALCulate<1|2>:MARKer<1 to 4>:COUNt ON | OFF
This command switches on or off the frequency counter at the marker position.
Example: ":CALC:MARK:COUN ON"
Features: *RST value: OFF
SCPI: device-specific
Mode: A
:CALCulate<1|2>:MARKer<1 to 4>:COUNt:RESolution 0.1 | 1 | 10 | 100 | 1000 | 10000 Hz
This command specifies the resolution of the frequency counter.
Example: ":CALC:MARK:COUN:RES 1kHz"
Features: *RST value: 1kHz
SCPI: device-specific
Mode: A
The numeric suffix in MARKer<1 to 4> is not significant.
:CALCulate<1|2>:MARKer<1 to 4>:COUNt:FREQuency?
This command queries the result of the frequency counter.
Example: ":CALC:MARK:COUN:FREQ?"
Features: *RST value: -
SCPI: device-specific
Modes: A, BTS, MS
This command is only a query and thus has no *RST value.
:CALCulate<1|2>:MARKer<1 to 4>:COUPled[:STATe] ON | OFF
This command switches the coupling of markers on or off.
Example: ":CALC:MARK:COUP ON"
Features: *RST value: OFF
SCPI: device-specific
Modes: VA
The numeric suffix in MARKer<1 to 4> is not significant.
:CALCulate<1|2>:MARKer<1 to 4>:LOEXclude ON | OFF
This command switches the local oscillator suppression on or off.
Example: ":CALC:MARK:LOEX ON"
Features: *RST value: OFF
SCPI: device-specific
Mode: A-F
The numeric suffixes 1|2 and 1 to 4 are not significant.