Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
150User Manual 1173.9286.02 ─ 15
8.15.1.3 STATus:QUEStionable:LIMit Register
The STATus:QUEStionable:LIMit register contains information about limit lines
and the results of a limit checks.
The number of LIMit registers depends on the number of measurement windows avail-
able in any application.
Bit no Meaning
0 LIMit 1 FAIL
This bit is set if limit line 1 is violated.
1 LIMit 2 FAIL
This bit is set if limit line 2 is violated.
2 LIMit 3 FAIL
This bit is set if limit line 3 is violated.
3 LIMit 4 FAIL
This bit is set if limit line 4 is violated.
4 LIMit 5 FAIL
This bit is set if limit line 5 is violated.
5 LIMit 6 FAIL
This bit is set if limit line 6 is violated.
6 LIMit 7 FAIL
This bit is set if limit line 7 is violated.
7 LIMit 8 FAIL
This bit is set if limit line 8 is violated.
8 Phase Noise LIMit FAIL
This bit is set if a limit of the phase noise limit line is violated.
9 to 14 Unavailable for phase noise measurements.
15 This bit is always 0.
8.15.1.4 STATus:QUEStionable:PNOise Register
The STATus:QUEStionable:PNOise register contains information about the status
of phase noise measurements.
Bit no Meaning
0 No trace are active
This bit is set if no trace is on.
1 SIGNal not found
This bit is set if no valid signal could be found.
2 VERify
This bit is set if signal verification has failed.
Using the Status Register