EasyManua.ls Logo

Rohde & Schwarz FSW-K40 - Page 150

Rohde & Schwarz FSW-K40
166 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
150User Manual 1173.9286.02 ─ 15
8.15.1.3 STATus:QUEStionable:LIMit Register
The STATus:QUEStionable:LIMit register contains information about limit lines
and the results of a limit checks.
The number of LIMit registers depends on the number of measurement windows avail-
able in any application.
Bit no Meaning
0 LIMit 1 FAIL
This bit is set if limit line 1 is violated.
1 LIMit 2 FAIL
This bit is set if limit line 2 is violated.
2 LIMit 3 FAIL
This bit is set if limit line 3 is violated.
3 LIMit 4 FAIL
This bit is set if limit line 4 is violated.
4 LIMit 5 FAIL
This bit is set if limit line 5 is violated.
5 LIMit 6 FAIL
This bit is set if limit line 6 is violated.
6 LIMit 7 FAIL
This bit is set if limit line 7 is violated.
7 LIMit 8 FAIL
This bit is set if limit line 8 is violated.
8 Phase Noise LIMit FAIL
This bit is set if a limit of the phase noise limit line is violated.
9 to 14 Unavailable for phase noise measurements.
15 This bit is always 0.
8.15.1.4 STATus:QUEStionable:PNOise Register
The STATus:QUEStionable:PNOise register contains information about the status
of phase noise measurements.
Bit no Meaning
0 No trace are active
This bit is set if no trace is on.
1 SIGNal not found
This bit is set if no valid signal could be found.
2 VERify
This bit is set if signal verification has failed.
Using the Status Register

Table of Contents

Related product manuals