Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
87User Manual 1173.9286.02 ─ 15
LAYout:ADD[:WINDow]? <WindowName>,<Direction>,<WindowType>
This command adds a window to the display in the active measurement channel.
This command is always used as a query so that you immediately obtain the name of
the new window as a result.
To replace an existing window, use the LAYout:REPLace[:WINDow] command.
Parameters:
<WindowName> String containing the name of the existing window the new win-
dow is inserted next to.
By default, the name of a window is the same as its index. To
determine the name and index of all active windows, use the
LAYout:CATalog[:WINDow]? query.
<Direction> LEFT | RIGHt | ABOVe | BELow
Direction the new window is added relative to the existing win-
dow.
<WindowType> text value
Type of result display (evaluation method) you want to add.
See the table below for available parameter values.
Return values:
<NewWindowName> When adding a new window, the command returns its name (by
default the same as its number) as a result.
Example:
LAY:ADD? '1',LEFT,MTAB
Result:
'2'
Adds a new window named '2' with a marker table to the left of
window 1.
Usage: Query only
Table 8-3: <WindowType> parameter values for Phase Noise application
Parameter value Window type
FDRift Frequency drift
MTABle Marker table
PNOise Phase noise diagram
RNOise Residual noise table
SNOise Spot noise table
SPECtrum Spectrum monitor
SPURs Spur list
SRESults Sweep result list
STABility Frequency and level stability indicator