Model 3095FB Multivariable Transmitter with Modbus Protocol
Modbus Protocol Guide
Page 55 of 79
0037 Coil R/W, WP Restore
Nonvola-
tile Data-
base
Restore Sensor Section 1
DP unit code
SP unit code
PT unit code
RTD present flag
User Entered PT Value
Lower Operating Limits
Upper Operating Limits
Inches H20 @ 60 F
psi
F
ON
60 F
Lower Range Limit
Upper Range Limit
0039 Coil R/W, WP Restore Trim Section 1
DP damping
SP damping
PT damping
Scale Integer Method
DP_y1
DP_y2
SP_y1
SP_y2
PT_y1
PT_y2
DP_x1
DP_x2
SP_x1
SP_x2
PT_x1
PT_x2
Maximum Integer
DP Scale Factor
DP Offset
SP Scale Factor
SP Offset
PT Scale Factor
PT Offset
.864
.864
.864
0
0
65534
0
65534
0
65534
-250
250
0
800
-40
400
65534
131
18
81
32768
148
26848
0041 Coil R/W, WP Restore
Nonvola-
tile Data-
base
Restore LCD Section 1
LCD Bit Mask
LCD Update Rate
0x0000 000f
3 seconds
Table 9-3 Restoring Nonvolatile Database