Plan of terminal connections for con::cube D-330-230
Plan of terminal connections for con::cube D-330-024
5.2 Connection of s::can Spectrometer Probe
An s::can spectrometer probe can be connected to the external connector for spectrometer probes, which is marked with no.4
inthegureofsection3.3.Ifanextensioncableisusedthetotallengthoftheprobe‘scableshouldnotbemorethan40m.
If necessary, a con::nect box (B-23-xxx) has to be used to connect a further spectrometer probe.
5.3 Connection of s::can ISE Probes and Sensors
Alternative or additional to the spectrometer probe it is possible to connect ISE probes, i::scan and / or s::can sensors via
thesysplugconnectorslocatedonthecon::cube.Theseconnectorsaremarkedwithno.5inthegureofsection3.3.Ifan
extension cable is used the total length of the probe’s cable should not be more than 40 m.
If necessary, the distribution box (B-41-HUB) has to be used to increase the number of plugs.