8.1.Mode Set
Click on the “MODE SET”’s pattern in Figure 19 to enter Mode Set screen.
Mode Set Screen
Figure 21
Figure 22
1 Run mode: selecting the run mode of test, Fix or Program mode.
Selecting Program Mode will display as Figure 22.
2 Run time: setting the running time for the FIX mode test, another way to set
the Run Time of FIX Mode. See 8.2 FIX Set
Prog. NO.: selecting the run program NO. of Program Mode.
3 Boot mode: selecting the restore status after power off during test.
Stop: after power off, restore to the SV of current running Program
or Fix Mode. But test is stop running, need to re-start by manual.
Cold: after power off, restore to the first segment of current running program,
or the SV of FIX Mode which was running before power off. And test is
auto-running.