Cubis MSE Operating Instructions 27
Getting Started
2. Push the cover of the below-cell weighing port to one side�
3. Carefully screw on the below-cell weighing hook�
3
Do not screw it in too tightly, as this could damage the thread
or the balance�
h
Install a draft protection shield�
t Attach the sample (e�g�, using a suspension wire) to the hook�
4. When weighing has been completed, unscrew the hook and return it to the clip�
t Close the cover of the below-cell weighing port�
Balances with a readability of ≤ 1 µg
t Remove both screws from beneath the weigh cell and detach the cover plate�
t Attach a wire, for example, to the sample and hang it on the notched hook�
h
Install a draft protection shield�
Balances with a Weighing Range of 20 kg or more:
t Use a suitable screwdriver to unscrew the cover plate from the bottom of the
balance�
t Request the hook directly from Sartorius� The hook is available as an accessory�
h
Install a draft protection shield�
The below-cell weighing port may not be opened or used on balances
used for legal metrology�
2
3