Operation
Cubis MSE Operating Instructions 31
Level the Balance (Q-Level)
It is essential for exact weighing results that the balance is absolutely level� The
front leveling feet can be used to level out small tilts in the floor� An integrated
sensor detects the alignment of the balance and triggers a warning message when
leveling is required� When “ " flashes on the display, you must level the balance�
Leveling with the inclination sensor:
Step Press key Display
1�
flashes on the display: flashes
Start leveling� * SCREW IN
2� Fully screw in both back leveling feet
(depends on the model) and confirm� * TURN
3� Directional arrow on the right of TURN:
Turn right leveling foot� TURN
Directional arrow on the left of TURN:
Turn left leveling foot� TURN
Directional arrow
:
Turn the respective leveling
foot to the right until the
directional arrow disappears�
Directional arrow : Turn the respective leveling
foot to the left until the
directional arrow disappears� SCREW OUT
4� Screw out both back leveling feet
(depends on the model) until they
touch the setup surface� * 0.0g
Leveling the balance with motorized leveling feet (optional):
t To start the leveling function, press the * key� flashes
AUT.LEV
y You can continue when the balance displays a weight value�
3
Each time the balance is leveled, the balance must be calibrated/adjusted
(see Calibrating/Adjusting the Balance)�
Turning the ionizer on/off (optional)
Purpose: The electrostatically charged weighing sample is neutralized as much as
possible in the air flow of the ionizer�
If your balance is equipped with an ionizer module, you can use the * key to
turn the ionizer on/off�
Preparation
t Set the intensity of the ionization in the menu�
Menu setting: extras: device: ionisat.: *Off, Soft intensity, Normal
intensity
or Strong intensity

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model