Application Programs
54 Cubis MSE Operating Instructions
Averaging (animal weighing)
Display symbol: V
Purpose
This application is used to determine the weights of unstable samples (e�g�, live
animals) or to determine weights under very unstable ambient conditions�
This requires an average to be calculated over several measurement cycles (also
referred to as “subweighing operations”)�
Changing the number of subweighing operations
Activate function: Press the S key
Select the desired number of measurements (1 to 100):
In increments of one: Press the S key briefly
In increments of ten: Press and hold the S key
The selected number of subweighings is saved to protected memory�
Preparation
§ Select the “Animal Weighing” application in the menu:
See “Configuration”
§ Configuring parameters:
Applic. Application programs
Anim.WG Animal weighing
Activty. Animal activity
Calm stable
o Active unstable
UNSTABL very unstable
Start
manual
o Autom. Automatic
o = Factory settings
Printout: Averaging
mDef 20 : Defined number of subweighing operations
x-Net + 410.1 g : Result of averaging

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model