Configuration (Operating Menu)
38 Cubis MSE Operating Instructions
Parameter Settings: Overview
ο = Factory setting = User-defined setting
Info about
Level 1 Level 2 Level 3 Level 4 Menu Level
1) Setup BAL.SCAL. AMBIENT conditions (adapt filter) V.Stable Very stable 1� 1�1� 1
Balance ο STABLE 1� 1�1� 2
parameters UNSTABL 1� 1�1� 3
V.UNSTBL. Very unstable 1� 1�1� 4
App.Filt. Application filter ο FINAL.RD. Final readout mode 1� 1�2� 1
Filling Filling mode 1� 1�2� 2
REDUC. Reduced 1� 1�2� 3
OFF 1� 1�2� 4
Stab.Rng. Stability range Max. ACC. Maximum accuracy 1� 1�3� 1
V.ACC. 1� 1�3� 2
ο ACC. 1� 1�3� 3
FAST 1� 1�3� 4
V.FAST. 1� 1�3� 5
Max.FAST
1
) 1� 1�3� 6
ST.DEL� Delay No 1� 1�4� 1
ο Short 1� 1�4� 2
AVERG. 1� 1�4� 3
LONG 1� 1�4� 4
Tare
1
) WIOSTB.W/o stability 1� 1�5� 1
ο WISTB.After stability 1� 1�5� 2
ATSTAB. At stability 1� 1�5� 3
Aut.zero Off 1� 1�6� 1
Auto zero ο on 1� 1� 6� 2
WT.UNIT For list of units, see 1� 1�7� 1
Basic weight unit “Toggling Between Weight Units” to
1� 1�7�24
DISP.DIG.
1
) ο All 1� 1�8� 1
Display accuracy lp.on.off (last digit after load change) 1� 1�8� 2
DIV.1 1 interval 1� 1�8� 6
Minus 1 Reduced by 1 digit 1� 1� 8� 7
Cal./ADJ. EXT.CAL. External calibration with standard weight
2
) 1� 1�9� 1
Function of C key e.cal.usr. External calibr�/adjustment with user-defined weight
2
) 1� 1�9� 3
ο Cal.Int. Internal calibration 1� 1�9� 4
Int.lin. Internal linearization (on analytical balances only) 1� 1�9� 5
Set.Prel. Setting the Preload 1� 1�9� 8
Del.Prel. Clearing the Preload 1� 1�9� 9
blocked C Blocked 1� 1�9�10
Select 1� 1�9�12
set.ext.w. Determine ext� calibration weight for e.cal.usr. 1� 1�9�17
Cal.adj.ON ο Sequence adjustment 1� 1�10� 1
CAL.ADJ. Adjustment as needed 1� 1�10� 2
Zero rng. Zero range 1 perc.ent 1� 1�11� 1
ο 2 perc.ent 1� 1�11� 2
INIT.ZERO Zero at Power On ο default Factory setting 1� 1�12� 1
2 perc.ent 1� 1�12� 2
On Tare Tare/zero at power on ο ON 1� 1�13� 1
OFF 1� 1�13� 2
CYC.rate Output rate ο normal 1� 1�14� 1
highvar. 1� 1�14� 2
slow 1� 1�14� 3
medium 1� 1�14� 4
fast 1� 1�14� 5
veryfast 1� 1�14� 6
maximum 1� 1�14� 7
ISOcal Autom� calibration/adjustment Off 1� 1�15� 1
Note 1� 1�15� 2
ο On 1� 1�15� 3
To.w.linearization (for analytical balances only) 1� 1�15� 4
EXT.Cal. External calibration
1
) ο Free 1� 1�16� 1
Locked (blocked) 1� 1�16� 2
cal.unit
1
) ο Gram 1� 1�17� 1
for calibration weight Kilogr. Kilograms 1� 1�17� 2
USERDEF. (factory setting: pound) 1� 1�17� 4
GEN.SERV. men.Reset Menu reset Yes Restore fcty� settings 1� 9� 1� 1
General servicing (Factory Settings) ο no Do not restore settings 1� 9� 1� 2
Standard
1
) 1� 9� 1� 3
Verifiable
1
) 1� 9� 1� 4
1
) Not available on balances verified for use in legal metrology
2
) Not available on balances for use in legal metrology of accuracy class K

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model

Related product manuals