ISO/GLP-compliant Printout
62  Cubis MSE Operating Instructions
ISO/GLP-compliant Printout
Features
You can have the device information, ID and the current date 
printed before (GLP header) and after (GLP footer) the values 
from the weighing series� These parameters include: 
GLP Header:
– Date
–  Time at the start of a weighing series
–  Balance manufacturer
–  Balance model
–  Model serial number
–  Software version number
–  ID number (of weighing series)
 GLP footer:
– Date
–  Time at the end of a weighing series
–  Field for signature
Configuration
t  Set the following menu items  
(Setup mode, see “Configuration”):
–  Line format for printout with ID –  
22 characters (factory setting): print - Prnt.para. - 
format - 22 chars
–  ISO/GLP-compliant printout or record for calibration/
adjustment only:  
data out- Prnt.para. - Glp - Cal.|Adj. 
 or  
ISO/GLP-compliant printout or record always on:  
data out- Prnt.para. - Glp -always on
–  Time formats: 
data out- Prnt.para. - time - 24h 
  or   
data out- Prnt.para. - time -12h  
with “Ahh:mm” or “Phh:mm”  
–  Date formats: 
data out- Prnt.para. - Date - dd.mmm.yy  
  or   
data out- Prnt.para. - Date - mmm.dd.yy 
!  No ISO|GLP-compliant record is output if any of the 
following settings are configured: 
data out- prnt.para.-format -16 chars 
 Operation 
t  To print the header and the first measurement value:  
Press the P key�
y  The header is included with the 1st printout�
t  To print the header and reference data automatically 
when an application program is active:  
Press the V key
t  Print the footer/Exit the application:
Application Functional process
Density Determination,  
Net-total Formulation, 
Totalizing
Print footer and quit  
application program:  
1+ Press the F key
Counting, Weighing in 
Percent Calculation,  
Animal Weighing
1)  Print footer: press F
2)  Quit application program: 
press F
The ISO/GLP-compliant printout can have the following lines:
--------------------  Dash line
17-Aug-2008    10:15  Date/Time (start of measurement)
      SARTORIUS  Balance manufacturer
Mod.        MSE8201S  Balance type
Ser. no.    10105355  Balance serial number
Ver. no.    00-39-04  Software version
ID           690 923  ID no�
--------------------  Dash line
L ID  Weighing series no� (Lot ID)
nRef          10 pcs  Counting: Reference sample quantity
wRef       21.14 g  Counting: Reference weight
Qnt   +      567 pcs  Counting results
--------------------  Dash line
17-Aug-2008    10:20  Date/Time (end of measurement)
Name:  Field for signature 
  Space line
--------------------  Dash line
The ISO/GLP-compliant printout for external calibration/adjustment:
--------------------  Dash line
17-Aug-2008    10:30  Date/Time (start of measurement)
      SARTORIUS  Balance manufacturer 
Mod.        MSE8201S  Balance type
Ser. no.    10105352  Balance serial number
Ver. no.    00-39-04  Software version
ID           690 923  ID no�
--------------------  Dash line
Cal. Ext. Test  Calibration/adjustment mode
Set   +   5000.0 g  Calibration weight value
Diff. +      0.2 g  Difference after calibration
Cal. Ext. Complete   Confirmation of completed calibration 
process
Diff.        0.0 g   Difference between current and target 
values after calibration
Level ok  Balance level
--------------------  Dash line
17-Aug-2008    10:32  Date/Time (end of measurement)
Name:  Field for signature 
  Space line
--------------------  Dash line