72 Cubis MSE Operating Instructions
Data Interfaces
Overview of SBI Commands
Format Command Action/Function Comments
1 ESC P Print to the interface where the prompt originated Corresp� to menu, with/
without stability
1 ESC T “TARE” key taring and zeroing
1 ESC K Filter “Very stable conditions”
1 ESC L Filter “Stable conditions”
1 ESC M Filter “Unstable conditions”
1 ESC N Filter “Very unstable conditions”
1 ESC O Block keys
1 ESC Q Acoustic signal
1 ESC R Unblock keys
1 ESC S Restart
1 ESC Z Internal calibration/adjustment Depending on menu,
1/2step increments
1 ESC U Tare
1 ESC V Zero
1 ESC W Ext� Adjustment with default weight Depending on menu,
1/2step increments
2 ESC f0_ S key
2 ESC f1_ Start calibration
2 ESC f2_ V key
2 ESC f5_ Left draft shield key (closing and opening as learned or default) Only if available
2 ESC f6_ Right draft shield key (closing and opening as learned or default) Only if available
2 ESC kP_ Print as with “PRINT” key (e�g�, to several interfaces)
2 ESC m0_ Ionizer status Only if available
2 ESC m1_ Ionizer on, with preset time Only if available
2 ESC m2_ Ionizer off Only if available
2 ESC s3_ F key: Back, exit, cancel
Models with the Models with the
analytical draft shield: rotational draft shield
2 ESC w0_ Draft shield status Draft shield status Only if available
2 ESC w1_ Open left door Open draft shield 100% to the left Only if available
2 ESC w2_ Close all doors Close draft shield Only if available
2 ESC w3_ Open upper door Open draft shield up to position saved Only if available
2 ESC w4_ Open right door Open draft shield door 100% to the right Only if available
2 ESC w5_ Open left and upper doors Only if available
2 ESC w6_ Open left and right doors Only if available
2 ESC w7_ Open right and upper door Only if available
2 ESC w8_ Open all doors Only if available
2 ESC x1_ Print weigher type
2 ESC x2_ Output serial number of
2 ESC x3_ Print balance software version
2 ESC s0_ Press and hold the S key

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model