Technical Specifications
Model-specific Data
Microbalances 0�001mg
Model MSE6.6S MSE6.6S-F MSE3.6P
Readability mg 0�001 0�001 0�001/0�002/0�005
Weighing capacity g 6�1 6�1 1�1/2�1/3�1
Tare range (subtractive) g -6�1 -6�1 -3�1
Repeatability <±mg 0�001 0�001 0�003/0�004/0�005
Linearity <±mg 0�004 0�004 0�004
Corner load (test load [g])*) µg 4 (2 g) 4 (2 g) 5 (1 g)
Optimal starting point of the measuring range**)
mg 0�82 0�82
Sensitivity drift between +10 to +30°C ±ppm/K 1 1 1
Typical stabilization time s < 5 < 5 < 5
Typical measurement time s < 8 < 8 < 8
External standard calibration value g 5 (E2) 5 (E2) 3 (E2)
(of at least accuracy class���)
Display result
(depending on the set filter level) 0�1 – 0�4 0�1 – 0�4 0�1 – 0�4
Weighing pan size d mm 30 50/30*) 30
Weighing chamber height mm 70 15 70
Protection Protected against dust and water
Ultramicrobalances 0�0001mg
Model MSE2.7S MSE2.7S-F
Readability mg 0�0001 0�0001
Weighing capacity g 2�1 2�1
Tare range (subtractive) g –2�1 –2�1
Repeatability <±mg 0�00025 0�00025
Linearity <±mg 0�0009 0�0009
Corner load (test load [g])*) µg 0�5 (1 g) 0�5 (1 g)
Optimal starting point of the measuring range**)
mg 0�082
Sensitivity drift between +10 to +30°C ±ppm/K 1 1
Typical stabilization time s < 7
Typical measurement time s < 10
External standard calibration value g 2 (E2) 2 (E2)
(min� accuracy class)
Display result
(depending on the set filter level) 0�1 0�4 0�1 0�4
Weighing pan size d mm 20 50/20*
)
Weighing chamber height mm 70 15
Protection Protected against dust and water
*) = Standard pan
**) = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�
88 Cubis MSE Operating Instructions

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model

Related product manuals