Cubis MSE Operating Instructions 99
Technical Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Precision balances
Modelle MSE12201S- MSE8201S- MSE5201S- MSE70200S- MSE36200S-
0CE 0CE 0CE 0CE 0CE
Accuracy class* K K K K K
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 100 100 100 1,000 1,000
Weighing capacity max* g 12,200 8,200 5,200 70,200 36,200
Calibration value e* g 1 1 1 10 1
Min� load min* g 5 5 5 50 50
Tare equalization range (subtractive) < 100% from max� weighing capacity
Application range according to DIR* g 512,200 58,200 55,200 5070,200 5036,200
Optimal starting point
of the measuring range** g 82 82 82 820 820
Typical stabilization time s < 0�8 < 0�8 < 0�8 < 1 < 1
Typical measurement time s < 1 < 1 < 1 < 1�2 < 1�2
Application range (temperature) With “isoCAL” function: +5 to +40 °C | Without “isoCAL” function: +10 to +30 °C
Display result
(depending on the set filter level) 0�1 – 0�4
Weighing pan size (W + D) mm 206 + 206 206 + 206 400 + 300 400 + 300 400 + 300
IP protection IP54
* DIR = Directive 2014/31/EU on non-automatic weighing instruments used within the European Economic Area
** = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model

Related product manuals