93
3+ soft key o Return to the “Device Parameters” menu..
Soft key q repeatedly Select the Operating parameters menu.
6+ soft key q, soft key O Select and open the Display geogr. data submenu.
If not already deactivated (“Off” marked by o), deactivate the display of geographical
data
(“Off”; factory setting). To do this, press the Q soft key to move the highlight bar to
“Off”
and press the l soft key to confirm. The circle (o) now marks the “Off” setting.
Important Note:
When you activate the “Factory Settings for All Parameters” menu item at the highest
menu level, all operating parameters are returned to the factory default settings.
This also applies to any customer-specific menu settings that have been configured.
For details, see “Restoring All Operating Parameters to Factory Settings” in this chapter.
§ Move the menu access switch from the “Accessible” to the “Closed” position:
§ If necessary, remove the cap that covers the menu access switch on the left-hand side of
the back of the indicator.
§ Move the menu access switch to the left (“Closed” position).
See also “Calibration and Adjustment” in the chapter entitled “Operation,” or refer to the
service manual.
§ Replace the protective cap over the menu access switch.
ee Restart the scale: Turn the indicator off and then on again.
The Sartorius logo is displayed briefly, after which the device is in normal weighing mode.
If the A/D converter was configured with a “Verifiable” data record,
the lines for display of metrological data (lines 1 and 2) show the data valid
for use in legal metrology when the menu access switch is closed.
For details, see the section entitled “Use in Legal Metrology” in this chapter.