Practical Example: Simple Counting
The QC Counting Scale is set at the factory so that you can begin counting without
changing any settings.
Step/Key Display Data Output
Place the container
on the scale, press
=/z (14) to tare
0.0 gcs
Place the appropriate
number of pieces (e.g.10)
in the container
x.x gcs
Press u/w (12),
the counting function starts.
10 pcs
Place any number of pieces in the container; the display shows the number
of pieces (“pcs”).
Note:
When –UPD– is displayed, the scale automatically re-calculates the reference
weight from the piece count on the scale and the current weight; this increases
the accuracy.
If you want to change the factory set reference sample quantity, press
g/W key (13) to select one of the pre-set numbers (5,10, 20, 50, 100)
or user-defined entry (factory setting 999*).
* = user-defined piece count to be stored in memory.