OM-610-1002-1 Operator’s Manual Rev 37
PDS-2010 SECTION 3 EQUIPMENT SPECIFICATION page 7
SECTION 3 EQUIPMENT SPECIFICATION
3.1 PDS-2010
Dimensions Width 19.5 in (49.5cm)
(Nominal) Height 50.5 in (128.3cm)
Depth 23.75 in (60.4cm)
Operating An ambient temp or 60–80°F is desired. Temperatures above 90°F will decrease
Environment coating efficiency and may cause electric equipment to shut down.
Relative Humidity should be kept between 5%–95%.
Weight 170 lb. (77 kg)
Chamber Size 12 inch dia. X 12 inch high (8.5”/31.5cm Dia. X 11.19”/28.4cm H, useable size)
with other chamber designs available
Power 110/220 VAC, single phase, 50-60 Hz, 20/12 A
(Specified at time of purchase)
Dimer Capacity 125 grams
Vacuum Pump 8.5 CFM, two-stage, direct drive
Exhaust Approx. 8.5 CFM @ 0.1 Inch Water (14.5 M
3
/H)
Cold Trap Optional Mechanical Chiller, LN
2
(connect with 3/8” O.D. tube), or Manual Fill
Probe
Controls Computerized sequencing using microprocessor temperature and process
controllers. Automatic and semi-automatic operation with fault condition and
end-of-run monitoring for controlled shutdown sequence.
3.2 TUMBLER OPTION
Dimensions Attaches to Optional Chamber with Removable Lid.
No additional floor space required
Weight 12 lb. (5.4 kg)
Canister Size 6 inch dia. x 6 inch long.
Power 0–130 VDC, 0.3A (Supplied by Labcoter™)
Load Capacity 68 Cubic Inches (1.1L) [At 40% fill]
Controls Controlled from Labcoter™