3.18
component 2
concentration;
Max5 bytes Concentration for component 2 as defined
for measure display. (if fitted)
component 2 units; Max 3 bytes Units defined for component 2. (if fitted)
component 2 alarm
status;
4 bytes One byte for each alarm. Set to alarm
number (1,2,3 or 4) when alarm is raised,
<space> when alarm not raised. (if fitted)
component 2
Autocalibration
status;
1 byte or
15 bytes
Indicates Autocalibration phase.
0 = Not in Autocalibration l
1 = In span preflush
2 = In zero cal
3 = zero corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
zero before calibration and the zero after
calibration respectively.
4 = In span cal
5 = span corrected. This is then followed by
two additional values, which are separated
by commas, each 6 bytes that represent the
span before calibration and the span after
calibration respectively.
6 = In post flush
<CR><LF> 2 bytes Carriage return and Line Feed characters
(ASCII codes 13 and 10 respectively).
component 3 formula; Max 6 bytes Chemical formula defined for component 3.
(if fitted)
component 3
concentration;
Max 5 bytes Concentration for component 3 as defined
for measure display. (if fitted)
component 3 units; Max 3 bytes Units defined for component 3. (if fitted)
component 3 alarm
status;
4 bytes One byte for each alarm. Set to alarm
number (1,2,3 or 4) when alarm is raised,
<space> when alarm not raised. (if fitted)
Table 3.6 – Serial Data Format