3.1
SECTION 3 INSTALLATION – ELECTRICAL
3.1 Electrical Power Connections
For convenience the gas struts may be removed during installation by carefully levering
them off. Ensure that the cable used is suited to the installation environment, and is the
correct rating. Ensure that the fuse is suited to current rating of the analyser and that
all glands and cable connections are secured firmly.
The mains power connection to this instrument is made via an electrical filter unit
attached to the outside of the case. The power cable into this unit does not need to be
screened.
WARNINGS
• Lethal voltage: mains AC power supplies are potentially lethal. The
installer must ensure that the power supply has been isolated before
commencing installation.
• The installer must be satisfied that the 2500 analyser installation
conforms to the relevant safety requirements, national electrical code
and any other local regulations, and that the installation is safe for any
extremes of conditions which may be experienced in the operating
environment of the analyser.
WARNING
The following symbols are used on the warning label fitted on the inside of the
Source End Enclosure Cover.
CAUTION, ELECTRIC SHOCK HAZARD
CAUTION, CONSULT MANUAL