1.6
.
Sample Temperature Compensation
A type 'K' thermocouple in stainless steel sheath is supplied, and a continuous
temperature measurement is made by the 2500 over the range -50 to 200°C (-58 to
392°F). During calibration in the factory, an empirical sample temperature/span
concentration relationship is determined and stored in memory to calibrate the
compensation. Suitable for gas or liquid samples, and for safe and hazardous area
installations. Not available with heated cell.
1.4.5 Application Options
A range of standard and special application options is available with the 2500 and it will
have been supplied configured with one particular option to make the customer-
specified measurements. Full details of the application configuration (i.e. optical
materials, operating wavelengths, cell path length, wetted materials etc.) will be found
on the Product Specification document in the front of this manual.
WARNING
Standard transducers are not suitable for oxygen service.