High: high brightness
•
Medium: medium brightness (factory setting)
•
Off: Display off
When the display is switched off, it must be switched on again via SOPAS ET or IO-Link.
7.6.6 Device > Factory reset (reset to factory settings)
The device can be reset to the factory settings.
During the reset process, the functions of the device are temporarily unavailable.
7.6.7 Device > Laser control (switch measuring laser on/off)
The sender (measuring laser) can be switched off. No measurement is possible when
the sender is switched off. The sender can also be switched on (On) and off (Off) via the
IN1 input.
7.6.8 Interface > Output Q1/Q2 > Active state (switching point logic)
The switching point logic describes the relationship between the output state (active or
inactive) and the potential applied to the digital output (high or low).
Settings (depending on output mode)
•
LOW:
Output potential for active output state: low electrical voltage
Output potential for inactive output state: high electrical voltage
•
HIGH (factory default):
Output potential for active output state: high electrical voltage
Output potential for inactive output state: low electrical voltage
7.6.9 Interface, Hysteresis
The hysteresis is the distance difference between the switch-on and switch-off point.
If the measured distance fluctuates around the set switching point, the hysteresis is
necessary for stable switching behavior.
To achieve a more precise switching behavior, set a smaller value for the hysteresis. To
achieve more stable switching, set a larger value for the hysteresis.
The parameter is set in the selected output mode menu.
7.6.10 Interface > Output Q1
The Q1 output is a digital output. In addition, the output serves as a communication
line for bidirectional data transmission when using the IO-Link interface.
The Q1 output offers different output modes. If an output mode is selected, the
required settings can be taught in or set manually.
•
IO-Link
•
Single-point
•
Window (factory setting)
•
SP +/- offset
7.6.10.1 Setting the SP1/SP2 switching point
A switching point can be taught in (Teach in SP) or set manually (Edit SP). The SP2
switching point is only available in Window (switching window) mode.
If using a switching window, ensure that identical distance values are not taught in for
SP1 and SP2.