OPERATING INSTRUCTIONS | LFP INOX 8019921/15H0 / 2019-09-30 | SICK AG
Irrtümer und Änderungen vorbehalten
22
7 COMMISSIONING
Commissioning
1. Mount the sensor as appropriate to the installation conditions, see “5 Mounting”.
2. Log in to expert mode, see “8.4.1 Expert mode”.
3. Select measuring mode.
Access the EXPRT-CONFIG-MeasMd menu using the arrow and Set pushbut-
tons.
HiSpd: max. length = 2,005 mm, response time < 400 ms
HiAcc: max. length = 6,005 mm, response time < 2,800 ms, more stable mea-
sured values, recommended for liquids with low DK values and where TrsHld
is < 70.
4. Teach in static sources of interference in the tank.
Static sources of interference in the tank generated by pipes, struts, cou-
plings, or cleaning balls are taught into the system as standard.
Access the EXPRT-CONFIG-CalRng menu using the arrow and Set pushbuttons.
Where the following applies:
Teach-in depth starts from the LFP process connection
The teach-in depth should cover all interference signals
Maximum teach-in depth (recommended) = probe length
Set the value range: 95 mm … 6,005 mm
If the tank cannot be emptied completely, the CalRng teach-in depth must be
adapted accordingly.
The level must be at least 200 mm under the CalLen and/or below the end of
the probe.
5. Run the AutCal function.
Access the AutCal menu using the arrow and Set pushbuttons.
The following applies: The probe must not be covered with liquid in the CalRng
set in step 4 (teach-in depth + 200 mm).
Use the Set pushbutton to conrm the AutCal menu item and then use it again
to conrm the Ok? prompt.
The AutCal function is conrmed with !CalOK.
6. Analyze the signal quality.
The signal quality can be analyzed when the device is installed,
see “8.4.9 Evaluating signal quality”.
In the event of problems:
Reduce the value in the EXPRT-CONFIG-TrsHld menu.
Set parameter to HiAcc in the EXPRT-CONFIG-MeasMd menu.
Switch on lters in the Set lters menu.
Reduce the parameter in the EXPRT-CONFIG-MaxCol menu.
7. Congure the lter (see “8.4.2 Filtering measured values).
8. Run maximum change of level / plausibility check
(see “8.4.2 Filtering measured values”).
9. Congure the outputs (see “8.2 Conguring the switching outputs”).