OPERATING INSTRUCTIONS | LFP INOX 8019921/15H0 / 2019-09-30 | SICK AG
Irrtümer und Änderungen vorbehalten
48
10 OVERVIEW OF PARAMETERS
Parameter Description
Reset Resets the values for CalEmp and CalMed.
Probe Probe settings.
Length See “8.4.7 Conguring the probe length”.
CblLen See “8.4.10 Changing the coaxial cable length”.
Type Choose either Rod (probe rod) or Rope (cable probe).
Info Sensor information.
FrmVer Shows the rmware version.
SerNo Shows the serial number.
CalSta Displays the status of the container calibration.
Pulse = AutCal (calibrated), NoCal (not calibrated)
Foam = FomCal (calibrated), CalMis (not calibrated)
AppTag Measuring point name, can only be written via IO-Link.
DevTag Device name, can only be written via IO-Link.
SigQua Parameter describes the quality of the measuring signal.
SigQa1 See “8.4.9 Evaluating signal quality”.
SigQa2 See “8.4.9 Evaluating signal quality”.
SigQa3 See “8.4.9 Evaluating signal quality”.
PASSW See “8.4.1 Expert mode”.